J2140A Attenuator

KEY FEATURES:

  • Integrated unit includes 40dB, 20dB and 10dB
  • Cascade for greater attenuation
  • Improve noise floor or assure small signal measurement

Download the product spec sheet here.
Visit the "Support Page" for more downloads.

What's included:

J2140A Attenuator
User manual
1 Year Warranty

Questions? Don't hesitate to Contact Us

Subscribe to newsletter
Model Description Price Cart
J2140A J2140A Attenuator (50 Ohm) $250.00 Add to Cart

There are two common uses for attenuators when used in conjunction with the network analyzer. One is to attenuate the oscillator source signal. While this may seem odd, one of the most common errors in analyzer measurements is using a source signal that is too large. Even though the analyzer allows setting of the signal output amplitude, the lowest setting is often too high to allow an accurate small-signal measurement to be made. The correct amplitude is the smallest amplitude that exceeds the noise floor.

Attenuators are also useful for improving the dynamic range of the measurement. In some cases, as in measuring the open loop gain of an opamp as one example, the low frequency loop gain will be extremely large (100dB or more is not uncommon). Attenuating the output signal increases the effective range of the measurement.

J2140A Attenuator Frequency Sweep
Specifications
Characteristic Rating
Maximum Input Level +20dBm
Input Impedance 50 ohms
Output Impedance 50 ohms
3dB Frequency Range DC-50MHz
Maximum VSWR 1.3
Attenuation Accuracy 0.2dB
Impedance 50 Ohm
Absolute Maximum Voltage < 50VAC and 75VDC

Caution: To avoid equipment damage and/or severe injuries or death ensure that the absolute maximum ratings are observed and not exceeded at all times.